RIGOL User’s Guide
Publication Number: UGE22X00-1110
RP1018H
RP1018H High Voltage Probe
Sept. 2013
© 2013 RIGOL Technologies, Inc. All Rights Reserved.
-6-
1. Specification
Model RP 1018H
Division Ratio 1000:1
Input Resistance 200MΩ
Input Capacitor 1.5 PF
Max. Input: DC+AC peak k18 V CAT II
Max. Input: AC RMS. k12 V CAT II
Max. Loading Current 90uA
Bandwidth DC~150MHz
Rise Time s 2.4 n
Signal / Noise > k 60dB at 1 Hz ; > 50dB at 1MHz
DCV Accuracy 3% Full Range
ACV Accuracy k 3% at 1 Hz
Temp. Coefficient 200 PPM /
Compensation Range ~ 10PF 35PF
Safety -Meets EN61010 031 CAT II
Cable Length ± 2M 0.2M
Operation Temperature Range - 10 ~ 55
Humidity ) 85% RH or less (at 35
Storage - Temperature Range 20 ~ 70
Color: Handel/Body / Ye Black llow
Weight / Volume 460g /80(W) x 80(H) X 320(L) mm
2. Precautions Safety
Thank you for purchasing RP1018H high voltage probe. H igh voltage probe can
prevent high voltage testers from subjecting to unexpected shocks when high doing
voltage measurement . Before using the high voltage probe, the operator must work
read and fully understand the safety rules.
In addition, this high voltage probe only for the person who are trained, experienced,
or otherwise qualified to recognize hazardous situations and who trained in the safety
precautions that necessary to avoid possible injury when using such a device.
-7-
Do not work alone when working with high voltage circuits.
For your own safety, inspect the probes for cracks and frayed or broken leads before
each use. If defects are noted, DO NOT USE the probe.
Hands, shoes, floor and work bench must be dry. Avoid making measurements under
humid, damp or other environmental conditions that might affect the safety of the
measurement situation. If possible, always turn the high voltage source off before
connection or disconnection the probe.
The probe body should be kept clean and free of any conductive contamination. The
probe is for indoor use only.
3. Operation
1) Connect the divider probe common lead (alligator clip) to a good earth ground or
reliable ground.
2) Connect the BNC connector to the BNC input of your oscilloscope.
3) Select the desired range of your oscilloscope.
NOTE
Please turn the high voltage source off before making any connections.
4. Compensation Adjustment
Connect the probe to the oscilloscope, and input a 200Hz square wave to the probe tip.
After that adjust the oscilloscope controls to display a few cycles of the waveform.
Adjust the trimmer located in the BNC plug for a flat topped square wave.
Illustration of 200Hz square wave adjustment

Produktspecifikationer

Varumärke: Rigol
Kategori: Inte kategoriserad
Modell: RP1018H

Behöver du hjälp?

Om du behöver hjälp med Rigol RP1018H ställ en fråga nedan och andra användare kommer att svara dig




Inte kategoriserad Rigol Manualer

Rigol

Rigol NFP-3 Bruksanvisning

26 September 2024
Rigol

Rigol PVA7250 Bruksanvisning

25 September 2024
Rigol

Rigol PHA1150 Bruksanvisning

25 September 2024
Rigol

Rigol RM-DG Bruksanvisning

18 September 2024
Rigol

Rigol PVP2150 Bruksanvisning

18 September 2024
Rigol

Rigol PVP3150 Bruksanvisning

16 September 2024
Rigol

Rigol PHA2150 Bruksanvisning

16 September 2024
Rigol

Rigol RP7150S Bruksanvisning

16 September 2024
Rigol

Rigol PVP2350 Bruksanvisning

16 September 2024

Inte kategoriserad Manualer

Nyaste Inte kategoriserad Manualer