Instruction Manual
Model: Neb Tester
Portable Testing Device
for Nebulizer
www.rossmax.com
OBM_IB_NebTester_
EN_SW_ver2118
Manufacturer: Rossmax International Ltd.,
12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan
www.rossmax.com
Specications
Cautions and maintenance
1. This device should be used with nebulizer only.
2. Place the device and nebulizer on a at, horizontal surface while in use.
3. The device must be in a vertical position while in use.
4. Turn the regulator slowly to avoid possible damage.
5. Do not completely unscrew regulator stem.
6. Do not attempt to disassemble the device.
7. Avoid excessive vibration, physical impact and dropping.
8. Do not store the device in direct sunlight, high temperature or humidity.
9. Keep the device away from static electricity.
10. For the rst use, open the rubber vent-plug on the back side of pressure gauge and
then close it in order to balance the pressure after the transportation.
Note: Do not open the rubber vent-plug when the device is in a vertical position to avoid
leakages of oil material from the pressure gauge.
Power source No need
Operating temperature 0 ~ 54ºC
Storage temperature - 10 ~ 60ºC
Air ow scale 1 ~ 10 L/min
accuracy ± 4% of full scale
Pressure scale 0 ~ 60 PSI
accuracy ± 2% of full scale
Weight 1000 g
Dimension 128 (L)*120(W)*158 (H) mm

Rossmax Neb Tester is a portable and easy-to-use device that allows the user to quickly
check the performance of the pump of a compressor nebulizer. Power source is not
required for Neb Tester during operation and the test result is easy to read and interpret.
Neb Tester consists of an oil pressure gauge, ow meter, air tube and stainless steel stand.
This device is used to simulate the working conditions of the pump when the nebulizer is
in operation with a medication cup, and you are able to check:
1. Maximum (free) air ow
2. Operational air ow at a particular pressure
3. Maximum air pressure
Parts name/location
Introduction
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
Flow meter
Regulator
Pressure
gauge
Air tube
Insert air tube into connectorPlace the device in a
horizontal position
Open the rubber vent-plug
and then close it after
Back view
Installation
Operation procedures
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
1. Check maximum (free) air ow
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Adjust air ow and set the level of the pressure at 0 PSI by turning the
regulator of ow meter.
Step 6: Check the air ow rate.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
2. Check operational air ow at a particular pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the pressure to a desired level by adjusting the regulator of the ow
meter.
Step 6: Check the air ow rate.
Pressure (PSI) Flow rate (L/min)
NA100 / NB500 (N1) 15.0 ≥ 3.5
NE100 / NF100/ NJ100 (N4) 15.0 ≥ 3.5
NK1000/ NB80 / NF80 (N1) 8.0 ≥ 3.0
NB60 / NI60/ NH60 / NF60 (N1) 8.0 ≥ 2.0
NL100 (N2) 10.5 ≥ 3.5
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
3. Check maximum air pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the air ow rate at 0 L/min by turning the regulator of ow meter
clockwise completely.
Step 6: Check the values on the pressure gauge.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
Air
compressor
connector
Rubber
vent-plug
Produktspecifikationer
Varumärke: | Rossmax |
Kategori: | Medicinsk |
Modell: | Neb Tester |
Vikt: | 34650 g |
Bredd: | 1138 mm |
Djup: | 700 mm |
Höjd: | 2315 mm |
Antal hyllor: | 1 schappen |
Maximal viktkapacitet: | 100 kg |
Ramfärg: | Zilver |
VESA montering: | Ja |
Panelmonteringsgränssnitt: | 200 x 200,300 x 300,400 x 200,400 x 400,400 x 600,800 x 400,800 x 600,1000 x 600 mm |
Tiltvinkelområde: | -10 - 5 ° |
Maximal skärmstorlek: | 100 " |
Hjulhjul: | Ja |
Garantiperiod: | 1 jaar |
lutningsjusteringar: | Ja |
Lastad: | Nee |
Hyllor: | Ja |
Färg topp: | Zilver |
Färg planka: | Zwart |
Behöver du hjälp?
Om du behöver hjälp med Rossmax Neb Tester ställ en fråga nedan och andra användare kommer att svara dig
Medicinsk Rossmax Manualer

12 September 2024

10 September 2024

8 September 2024

7 September 2024

4 September 2024

1 September 2024

1 September 2024

31 Augusti 2024

30 Augusti 2024

28 Augusti 2024
Medicinsk Manualer
- Medicinsk Pulsar
- Medicinsk Vermeiren
- Medicinsk Flowflex
- Medicinsk Medtronic
- Medicinsk Dittmann
- Medicinsk Hybernite
- Medicinsk Diley Dreams
- Medicinsk VQ OrthoCare
- Medicinsk Gima
- Medicinsk Rhino Horn
Nyaste Medicinsk Manualer

15 September 2024

14 September 2024

14 September 2024

14 September 2024

14 September 2024

14 September 2024

14 September 2024

Medtronic MiniMed one-press serter - MMT-7512 Bruksanvisning
13 September 2024
11 September 2024

11 September 2024