Instruction manual digital light meter
SAUTER SP 200K
Version 2.1
04/2023
GB
SP BA- -e-2321
PROFESSIONAL MEASURING
2 - -e- SP BA 2321
GB
SAUTER SP 200K
V. 2.1 04/2023
Instruction manual digital light meter
Table of contents:
1 General information ....................................................................................... 3
2 Functions ........................................................................................................ 3
3 Technical description .................................................................................... 3
4 Function keys and part names ..................................................................... 4
5 Inuencing variables of the spectral sensitivity ......................................... 5
6 Steps for commissioning .............................................................................. 5
7 Check and replace batteries ......................................................................... 6
8 Maintenance ................................................................................................... 6
9 Examples of recommended illuminance levels ........................................... 6
SP BA- -e- 3 2321
1 General information
This digital illuminance meter is a precision instrument for measuring the illuminance
in the work area. The sensor has a full cosine correction for oblique light incidence.
The instrument is compact, robust and easy to use due to its design. The light sensitive
component used in this light meter is a very stable silicon diode with a long lifetime.
2 Functions
The device measures the illumination from 0 to 200000 lux or from 0 to 20000 FC
(foot candle)
High accuracy and fast response
Peak-Hold function for the acquisition of peak values
Unit and number display for easy reading
Automatic zero setting
Non-standard light sources are automatically corrected
Short rise and fall times
also suitable for LED lighting
3 Technical description
Display: 3 ½ digit LCD
Measuring range: 200; .000; 20.000; 200.000 Lux
(20,000 lux reading value x 10, 200,000 lux reading value x 100)
20; 200; 2,000; 20,000 FC
(20,000 FC Reading value x 10)
1 FC= 10.76 Lux
Exceeded display range: Display of the highest Digit "1" appears on the
display
accuracy: ± 4% rdg ± 10 digit up to 20,000 lux / 2,000 FC
± 5% rdg ± 10 digit up to 200,000 lux / 20,000 FC
Calibrated with a standard incandescent lamp at a colour temperature of
2856K
Repeat precision: ± 2%
Temperature Characteristic value: ± 1%/°C
Measuring rate: 2 times/sec
Photosensor: Silicon photo diode with lter
Working temperature: 0°C to 40°C (32°F to 104°F)
Humidity: 0 to 70 %RH