Klein Tools CL810 Bruksanvisning

Läs nedan 📖 manual på svenska för Klein Tools CL810 (27 sidor) i kategorin Mätutrustning. Denna guide var användbar för 2 personer och betygsatt med 4.5 stjärnor i genomsnitt av 2 användare

Sida 1/27
INSTRUCTION MANUAL
ENGLISH
FRANÇAIS p. 33
ESPAÑOL pg. 17
600A AC/DC Auto-Ranging
Digital Clamp Meter
True RMS
Measurement
Technology
NON-CONTACT
VOLTAGE TESTER
LOW IMPEDANCE
DATA & RANGE HOLD
AUDIBLE CONTINUITY
DIODE TEST
CAPACITANCE & FREQUENCY
TRANSFLECTIVE
REVERSE-CONTRAST DISPLAY
LIGHTED DIAL
Ω
CL810
2
mIP40
-40° – 1832°F
(-40° – 1000°C)
5000573
CAT IV
600V
CAT III
1000V
2
GENERAL SPECIFICATIONS
Klein Tools CL810 is an automatically ranging true root mean square
(TRMS) digital clamp-meter that measures AC/DC current via the clamp,
measures AC/DC voltage, resistance, continuity, frequency, inrush,
capacitance, and tests diodes via test-leads, and temperature via a
thermocouple probe. It also features a Low Impedance (LoZ) mode for
identifying and eliminating ghost or stray voltages, transflective reverse-
contrast display, and lighted dial.
Environment: Indoor: Do not expose to moisture, rain or snow.
Operating Altitude: 6562 ft. (2000m)
Relative Humidity: <80% non-condensing
Operating Temp: ° ° ° °32 to 122 F (0 to 50 C)
Storage Temp: ° ° ° °14 to 140 F (-10 to 60 C)
Accuracy: Values stated at 65° to 83°F (18° to 28°C)
Temp Coefcient: ° °0.1 × (Quoted Accuracy) per C above 28 C
or below 18 C, corrections are required when ambient working °
temp is outside of Accuracy Temp range
Auto Power-Off (APO): After 5 minutes of inactivity (unless
disabled)
Dimensions: 9.06" × 3.47" × 1.65" (230 × 88 × 42 mm)
Weight: 13 oz. (369 g) including batteries
Calibration: Accurate for one year
Standards:
IEC EN 61010-1, 61010-2-032.
Conforms to: UL STD 61010-1, 61010-2-032.
Certied to: CSA STD C22.2 #61010-1, 61010-2-032.
Pollution degree: 2
Accuracy: ± (% of reading + # of least significant digits)
Drop Protection: 6.6 ft. (2m)
Safety Rating: CAT IV 600V, CAT III 1000V, Class 2,
Double insulation
CAT III: Measurement category III is applicable to test and
measuring circuits connected to the distribution part of the
building’s low-voltage MAINS installation.
CAT IV: Measurement category IV is applicable to test and
measuring circuits connected at the source of the building’s
low-voltage MAINS installation.
Electromagnetic Environment: IEC EN 61326-1. This
equipment meets requirements for use in basic and controlled
electromagnetic environments like residential properties,
business premises, and light-industrial locations.
Specifications subject to change.
ENGLISH
5000573
4
ELECTRICAL SPECIFICATIONS
Function Range Resolution Accuracy
Frequency
(Auto-
Ranging)
1.000Hz to 9.999Hz 0.001Hz
±(1.0% + 5 digits)
99.99Hz 0.01Hz
999.9Hz 0.1Hz
9.999kHz 1Hz
99.99kHz 10Hz
500.00kHz 100Hz
Sensitivity: >2V to 220V RMS
Maximum Input: 1000V AC RMS or 1000V DC
Duty Cycle
1.0% to 99.9%
0.1%
±(1.2% + 2 digits)
Pulse width: Frequency width: >100µs, <100ms 5Hz to 10kHz
Sensitivity: >2V to 220V RMS 1000V DC or 1000V AC RMSMaximum Input:
Inrush
Current 600.0A 100mA ±(3.0% + 8 digits)
Frequency Range: Detection Range: 50 to 60Hz >2A
OTHER MEASUREMENT APPLICATIONS
Maximum Input: 1000V AC RMS or 1000V DC
Diode Test: Max. 1.5mA, open circuit voltage ~3.2V DC
Audible signal <50Ω, test current <1.5mAContinuity Check:
4 samples per secondSampling Frequency:
Input impedence >3kΩ, Low Impedence (Low Z):
Max input 1000V AC RMS or 1000V DC RMS
Auto Power-Off: After ~5 minutes of inactivity
Overload: "OL" indicated on display, overload protection 1000V in all settings
"-" on display indicates negative polarityPolarity:
3-5/6 digit, 6000 Count LCDDisplay:
ENGLISH
WARNINGS - GENERAL
To ensure safe operation and service of the meter, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
Before each use verify meter operation by measuring a known voltage
or current.
Never use the meter on a circuit with voltages that exceed the category
based rating of this meter.
Do not use the meter during electrical storms or in wet weather.
Do not use the meter or test leads if they appear to be damaged.
Use only with CAT III 1000V/CAT IV 600V rated test leads or better.
Ensure meter leads are fully seated, and keep fingers away from the
metal probe contacts when making measurements.
Use caution when working with voltages above 25V AC RMS or 60V DC.
Such voltages pose a shock hazard.
To avoid false readings that can lead to electrical shock, replace batteries
when a low battery indicator appears.
Do not attempt to measure resistance or continuity on a live circuit.
Always adhere to local and national safety codes. Use personal protective
equipment to prevent shock and arc blast injury where hazardous live
conductors are exposed.
To avoid risk of electric shock, disconnect leads from any voltage source
before removing battery door.
To avoid risk of electric shock, do not operate meter while battery door is removed.
WARNINGS