Maxwell MX-25201 Bruksanvisning

Maxwell Multimeter MX-25201

Läs nedan 📖 manual på svenska för Maxwell MX-25201 (9 sidor) i kategorin Multimeter. Denna guide var användbar för 11 personer och betygsatt med 4.5 stjärnor i genomsnitt av 2 användare

Sida 1/9
U Mser anUal
Használati Utasítás
ManUal de Utilizare
Užívateľská prírUčka
DIGITAL MULTIMETER
DIGITÁLIS MULTIMÉTER
MULTIMETRU DIGITAL
DIGITÁLNY MULTIMETER
Product code / Termékkód / Cod produs / Kód produkta:
25201
EN
HU
RO
SK
i
VC890C+
EN User Manual
sUMMary
This broad-spectrum measurement device
enables you to measure DCV, ACV, DCA, ACA,
resistance, capacity, diode, transistor, continuity
and temperature. Ideal for using in laboratories,
factories and even at home.
s afety reMarks
This device complies to the IEC10101 standards.
Read the manual carefully before use.
Do not measure a higher value than the set
range.
To avoid electric shock make sure that the
measuring wires are intact.
Disconnect the measuring wires while
changing function.
Select the proper function and range, avoid
incorrect usage.
Do not use the device if the battery holder lid
and back cover are not in their places.
Do not measure voltage while in resistance
measuring mode.
Remove the wires and turn the device o
before replacing the battery or fuse.
Safety signs:
Dangerous voltage
Grounding
Double insulation
Low battery power
CHaraCteristiCs
Display LCD display
Max. value display: 4 digit automatic
polarity display
Low battery power
display
Operating
environment (0-40) °C, R. H. 80%.
Battery
9V X 1 (NEDA
1604/6F22 or same
type)
Dimensions 175 X 93 X 55 mm
Weight 400g (with battery)
t eCHniCal data
DCV
Range Accuracy Resolution
200mV
±0.5%
100μV
2V 1mV
20V 10mV
200V 100mV
1000V ±0.8% 1V
Input resistance: 5m Ω at mV range , other
ranges: 10 mΩ
ACV
Range Accuracy Resolution
2V
±0.8%
1mV
20V 10mV
200V 100mV
750V ±1.2% 1V
Input resistance: 10MΩ
Overload protection: 1000V DC or AC peak value
Frequency response : (40~200)Hz
DCA
Range Accuracy Resolution
200μA ±0.8% 100nA
20mA 10μA
200mA ±1.2% 100μA
20A ±2.0% 10mA
Max. inward current
20A (testing time
should be 10 sec
max)
Fuse 0,2A/250V and 20A /
250 V fast fuse
EN User Manual
DCV measuring
Connect the black measuring wire to the
„COM” connector, and the red one to the V/Ω”
connector.
Set the function switch to the proper DCV
setting, and connect the wires to the circuit.
Note:
If you are unsure about the measured voltage range, set
the function switch to a high range.
If the LCD shows „1 it means overload, and the range
should be set to a higher value.
ACV measuring
Connect the black measuring wire to the
„COM” connector, and the red one to the V/Ω”
connector.
Set the function switch to the proper ACV
setting and connect the wires to the circuit.
Note:
If you are unsure about the measured voltage range, set
the function switch to a high range.
If the LCD shows „1 it means overload, and the range
should be set to a higher value.
DCA measuring
Connect the black measuring wire to the
„COM” connector, and the red one to the „mA
connector (max. 200mA), or connect the red
wire to the „20A” connector (max. 20A).
Set the function switch to the proper DCA
setting and connect the wires to the circuit.
Note:
If you are unsure about the measured voltage range, set
the function switch to a high range.
If the LCD shows „1 it means overload, and the range
should be set to a higher value.
Max inward current 200mA or 20A (depending on the
placement of the red wire), in case of an exceeding
current the fuse will melt.
ACA measuring
Connect the black measuring wire to the
„COM” connector, and the red one to the V/Ω”
connector.
Set the function switch to the proper ACA
setting, and connect the wires to the circuit.
Note:
If you are unsure about the measured voltage range, set
the function switch to a high range.
If the LCD shows „1 it means overload, and the range
should be set to a higher value.
Max inward current 1000V.
Resistance measuring
Connect the black measuring wire to the
„COM” connector and the red one to the V/Ω
connector.
Set the function switch to the proper
resistance setting and connect the wires to the
measured resistor.
Note:
If the measured resistance value is over the chosen max
value the LCD will show „1” and the function switch
needs to be set to a higher range. It may take a few
seconds for the device to stabilize when measuring
above 1MΩ
Capacity measuring
Connect the wire to the V/Ω connector and
the red one to the „COM” connector.
Set the function switch to a proper capacity
setting and connect the wires to the measured
capacitor.
Note:
If you are unsure about the measured capacity set the
function switch to a high range.
If the LCD shows „1 it means overload and the function
switch needs to be set to a higher range.
To be safe, discharge the measured capacitor before
measuring.
Diode and continuity test
Connect the black measuring wire to the
„COM” connector, and the red one to the V/Ω”
connector (Note: the red wire's polarity is „+”)
Set the function switch to the position,
connect the wires to the measured diode.
Connect the wires to the diode's measuring
points, if the device beeps, the resistance is
lower than appr. (70±20)Ω
Temperature measuring
Connect the red wire to the mA, and the black
one to the "COM" connector. Touch the end
of the thermometer to the measured object
and the temperature value is displayed on the
screen in degrees Celsius.
hFE
Set the function switch to hFE.
Place the part into the transistor connector
depending on whether it is NPN or PNP type.
Automatic turn-o
If the device is not used for 15 minutes it turns
o automatically. To use it again, turn the
function switch.

Produktspecifikationer

Varumärke: Maxwell
Kategori: Multimeter
Modell: MX-25201
Färg på produkten: Rood
Bredd: 68 mm
Djup: 68 mm
Höjd: 112 mm
Kijkglas: Ja
Knivmaterial: Keramisch
Justerbar sliptyp: Ja

Behöver du hjälp?

Om du behöver hjälp med Maxwell MX-25201 ställ en fråga nedan och andra användare kommer att svara dig




Multimeter Maxwell Manualer

Multimeter Manualer

Nyaste Multimeter Manualer