PCE Instruments PCE-HPT 1 Bruksanvisning

Läs nedan 📖 manual på svenska för PCE Instruments PCE-HPT 1 (14 sidor) i kategorin Termometer. Denna guide var användbar för 5 personer och betygsatt med 4.5 stjärnor i genomsnitt av 2 användare

Sida 1/14
Manual
Thermometer
PCE-HPT 1
PCE Americas Inc.
711 Commerce Way
Suite 8
Jupiter
FL-33458
USA
From outside US: +1
Tel: (561) 320-9162
Fax: (561) 320-9176
info@pce-americas.com
www.pce-instruments.com/english
www.pce-instruments.com
PCE Instruments UK Ltd.
Units 12/13
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@industrial-needs.com
CONTENTS
TITLE PAGE
1. Introduction………………………………………………………………….……………. 1
2. Specifications……………………………………………………………………………... 1
3. Symbol Definition and Button Location………......................................................
2
4. Operation Instructions……………………….…………………..................................
3
4.1 Power-Up……………...........................................................................................3
4.2 Connection the Temperature Probe………............................ .............................. 3
4.3 Selecting the Temperature Scale……………..…………………………………….. 3
4.4 Data-Hold Operation…………………………………………………………………..3
4.5 Back light Operation………………………………………………………………….. 3
4.6 Relative Operation……………………………………………………………............ 3
4.7 Record Operation…………………………………………………………................. 3
4.8 MAX/MIN Operation………………………….……………………………................ 3
4.9 Auto Power Off………………………………………………………………………… 4
4.10 Low Battery Condition………………………………………..... ………………….. 4
5. Temperature Measurement……………………………………....... ………………….. 4
6. Setting the Time…………………………………………………………........................ 5
7. Changing the Logging Interval………………………………………………………… 5
8. Clearing Memory…………………………………………………………………………. 5
9. Temperature Meter Calibration Setup ……….……………………………………….. 6
10. Communicating with a PC…………………………………………………………….. 7
Testlink PCE-HPT 1 Software…………………………………………………………… 7
11. Power Preparation…………………………………………………………………….. 12
12. Maintenance .…….……………………………………………………………………. 12
13. Temperature Probe.…………………………………………………………….......... 12
PCE-HPT 1 THERMOMETER
1
1. INSTRUCTION
This instrument is a digital thermometer for use with platinum-type temperature sensor.
Temperature indication follows IEC751 temperature table for PT-type sensor.
2. SPECIFICATIONS
Numerical Display: 5 digital liquid crystal displays
Measurement Range: -100°C ~ 400°C; -148°F ~ 752°F
Resolution: 0.01°C; 0.02°F;
Sensor types:
Platinum resistance temperature sensor for pt-100 4 wires.
ALPHA=0.00385
Environmental:
Operating Temperature and Humidity:
0°C ~50°C (32°F ~ 122°F) ; 0 ~ 80% RH
Storage Temperature:
-10°C to 60°C (14°F ~ 140°F); 0 ~ 80% RH
Altitude up to 2000 meters.
Accuracy: at ( 23 ± 5°C )
Range Accuracy
-100°C ~ 400°C ±(0.05% reading + 0.1°C)
-148°F ~ 752°F ±(0.05% reading + 0.2°F)
Temperature Coefficient:
For ambient temperatures from 0°C ~ 18°C and 2C ~ 50°C, for each °C ambient below 18°C or
above 28°C add the following tolerance into the accuracy spec.
0.005% of reading + 0.01°C ( 0.005% of reading + 0.02°F )
Note:
The basic accuracy Specification does not include the error of the probe please refer to the
probe accuracy specification for additional details.
Sample Rate: 2 times per second
Dimension: 185 x 65 x 36 mm(meter only)
Weight: 360g Approx.
Power requirement:
9V Battery;
AC adapter: 9V DC(7~10V Max.)/20mA Min (Plug Diameter: 3.5mmx1.35mm)
Battery Life: Approx. 100hours
Standard Accessory:
Pt-100 Probe(class A), Instruction manual, 9V Battery, windows software, carrying case, USB
cable.
Optional Accessories: AC adapter, RS-232 Cable.

Produktspecifikationer

Varumärke: PCE Instruments
Kategori: Termometer
Modell: PCE-HPT 1

Behöver du hjälp?

Om du behöver hjälp med PCE Instruments PCE-HPT 1 ställ en fråga nedan och andra användare kommer att svara dig




Termometer PCE Instruments Manualer

Termometer Manualer

Nyaste Termometer Manualer